User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDC 5000 100 PT DATE: 08-Feb-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 17 NUMBER OF LINES: 108 STANDARD: GenRes RTD-100 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 ASK+ X 1.003 DISP Press UUT LOWR DISP button. 1.004 DISP ÿ Connect the RTD-100 and the UUT as follows: 1.004 DISP ÿ [27][91]1m RTD-100 TO UDC 5000 1.004 DISP ÿ OUTPUT V ê A HI ÄÄÄÄÄÄÄÄÄ INPUT 21 1.004 DISP ÿ OUTPUT V ê A LO ÄÄÄÄÄÄÄÄÄ INPUT 22 1.005 DISP Connect a short between INPUT 22 and 23. 1.006 HEAD {} 1.007 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.008 HEAD {º FULL SCALE AND LINEARITY TEST º} 1.009 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.010 HEAD {}ÍÍ FULL SCALE AND LINEARITY TEST ÍÍ 1.011 JMP 2.001 1.012 EVAL 2.001 STD GenRes RTD-100 2.002 DISP Set the RTD-100 dials to 025.154. 2.003 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 2.004 ACC -300.0øF 0.01% 2.005 MEME 2.006 MEMC øF 1.0U 3.001 DISP Set the RTD-100 dials to 048.388. 3.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 3.003 ACC -200.0øF 0.01% 3.004 MEME 3.005 MEMC øF 1.0U 4.001 DISP Set the RTD-100 dials to 070.956. 4.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 4.003 ACC -100.0øF 0.01% 4.004 MEME 4.005 MEMC øF 1.0U 5.001 DISP Set the RTD-100 dials to 082.010. 5.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 5.003 ACC -50.00øF 0.01% 5.004 MEME 5.005 MEMC øF 1.0U 6.001 DISP Set the RTD-100 dials to 093.007. 6.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 6.003 ACC 0.00øF 0.0093U 6.004 MEME 6.005 MEMC øF 1.0U 7.001 DISP Set the RTD-100 dials to 0(10)0.000. 7.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 7.003 ACC 32.00øF 0.01% 7.004 MEME 7.005 MEMC øF 1.0U 8.001 DISP Set the RTD-100 dials to 103.900. 8.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 8.003 ACC 50.00øF 0.01% 8.004 MEME 8.005 MEMC øF 1.0U 9.001 DISP Set the RTD-100 dials to 114.681. 9.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 9.003 ACC 100.0øF 0.01% 9.004 MEME 9.005 MEMC øF 1.0U 10.001 DISP Set the RTD-100 dials to 135.969. 10.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 10.003 ACC 200.0øF 0.01% 10.004 MEME 10.005 MEMC øF 1.0U 11.001 DISP Set the RTD-100 dials to 156.904. 11.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 11.003 ACC 300.0øF 0.01% 11.004 MEME 11.005 MEMC øF 1.0U 12.001 DISP Set the RTD-100 dials to 177.472. 12.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 12.003 ACC 400.0øF 0.01% 12.004 MEME 12.005 MEMC øF 1.0U 13.001 DISP Set the RTD-100 dials to 197.700. 13.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 13.003 ACC 500.0øF 0.01% 13.004 MEME 13.005 MEMC øF 1.0U 14.001 DISP Set the RTD-100 dials to 217.533. 14.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 14.003 ACC 600.0øF 0.01% 14.004 MEME 14.005 MEMC øF 1.0U 15.001 DISP Set the RTD-100 dials to 237.057. 15.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 15.003 ACC 700.0øF 0.01% 15.004 MEME 15.005 MEMC øF 1.0U 16.001 DISP Set the RTD-100 dials to 256.204. 16.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 16.003 ACC 800.0øF 0.01% 16.004 MEME 16.005 MEMC øF 1.0U 17.001 DISP Set the RTD-100 dials to 274.996. 17.002 MEMI Use keyboard to enter actual UUT reading in øFarenhite. 17.003 ACC 900.0øF 0.01% 17.004 MEME 17.005 MEMC øF 1.0U